PHILIPS M1026B ANESTHETIC GAS ANALYZER

Received Dec 3, 2019 · Event occurred Nov 8, 2019

Report 3007409280-2019-00121 · MDR key 9406819

Device

Generic name

Anesthetic Gas Module

Model number

FGPH1026B

Product problems

  • Electrical /Electronic Property Problem
  • Mechanical Jam
  • Electrical /Electronic Property Problem
  • Mechanical Jam

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

THE COMPLAINT DEVICE WAS RETURNED. DEVICE EVALUATION IDENTIFIED THAT THE DEVICE CAME IN WITH THE MICRON FILTER DISCONNECTED AND THE PUMP WAS BAD. THE PUMP WAS REPLACED. THE FAN, FAN FILTER, FLOW RATE, GAS CALIBRATION, LEAK, POWER ON, PRESSURE SENSOR, AND FINAL VISUAL INSPECTION WERE CHECKED/TESTED AND ALL PASSED. THE ROOT CAUSE WAS DETERMINED TO BE THE DEFECTIVE PUMP. THIS TYPE OF EVENT WILL CONTINUE TO BE MONITORED.

Description of Event or Problem

REPORTEDLY, POST REPAIR, THE DEVICE'S O2 ZERO FAILED AND AGM ZERO FAILED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.