BMT SPLINED KNEE STM V2 16X80 80
Received Dec 2, 2019 · Event occurred Apr 5, 2018
Report 0001825034-2019-05418 · MDR key 9397774
Device
Generic name
Prosthesis, Knee
Manufacturer
Zimmer Biomet, Inc.Model number
N/ACatalog number
148306
Lot number
731640
Product problems
- Loosening of Implant Not Related to Bone-Ingrowth
- Loosening of Implant Not Related to Bone-Ingrowth
Patient
Not reported
- Pain
- Limited Mobility Of The Implanted Joint
- Pain
- Limited Mobility Of The Implanted Joint
Narrative
Additional Manufacturer Narrative
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. UPDATED: B4, B5, G4, H1, H2, H3, H6, H10. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. CLINICAL NOTES FOUND THAT THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO FEMORAL LOOSENING; EVIDENCE OF SCALLOPING WHICH RESULTED IN PAIN AND DIFFICULTY AMBULATING. MEDICAL RECORDS WERE NOT PROVIDED. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
Description of Event or Problem
NO ADDITIONAL INFORMATION REPORTED.
Additional Manufacturer Narrative
(B)(4). CONCOMITANT MEDICAL PRODUCTS: VANGUARD SSK 360 LEFT FEMUR CATALOG 185288 LOT 2823216, VANGUARD 360 FEMORAL AUGMENT CATALOG 185428 LOT 115580, VANGUARD 360 FEMORAL AUGMENT CATALOG 185428 LOT 354980, BIOMET 360 OFFSET ADAPTER CATALOG 185211 LOT 497570, BIOMET 360 TIBIAL TRAY CATALOG 185206 LOT 184200, BIOMET SPLINED KNEE STEM CATALOG 148306 LOT 277800, BIOMET 360 OFFSET ADAPTER CATALOG 185211 LOT 497570, VANGUARD SSK TIBIAL BEARING CATALOG 185104 LOT 985950, UNKNOWN COBALT BONE CEMENT. CUSTOMER HAS NOT INDICATED WHETHER THE PRODUCT WILL BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THE PATIENT UNDERWENT A LEFT KNEE REVISION ARTHROPLASTY APPROXIMATELY THREE YEARS POST IMPLANTATION DUE TO PAIN AND LOOSENING OF THE FEMORAL COMPONENTS. NO ADDITIONAL INFORMATION IS AVAILABLE.