inforMED
MalfunctionJIL

COMBUR-7-TEST 100 STR.

Received Oct 22, 2019 · Event occurred Sep 25, 2019

Report 1823260-2019-03813 · MDR key 9218270

Add to watch list
Product code (JIL)
Manufacturer (Roche Diagnostics International Ltd.)
Model (NA)

Device

Generic name

Urine Test Strips

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

11008552191

Lot number

38055002

Product problems

  • False Negative Result
  • False Negative Result

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE CUSTOMER RETURNED ONE VIAL OF COMBUR-7-TEST 100 STR OF THE AFFECTED LOT FOR INVESTIGATION. THE TEST STRIPS SHOWED NO ABNORMALITIES. THE CUSTOMER MATERIAL AND RETENTION MATERIAL WERE VISUALLY CHECKED AND CHECKED WITH NEGATIVE URINE AND WITH AN ERYTHROCYTE DILUTION SERIES AND A NITRITE DILUTION SERIES. NEITHER MATERIALS SHOWED ABNORMALITIES, AND BOTH FULFILL REQUIREMENTS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative

THE EVENT OCCURRED IN (B)(6).

Description of Event or Problem

THE INITIAL REPORTER COMPLAINED OF FALSE NEGATIVE NITRITE RESULTS FOR 1 PATIENT URINE SAMPLE TESTED WITH COMBUR-7 URINE TEST STRIPS. THE NITRITE RESULT WAS NEGATIVE. THE PHYSICIAN'S RESULT WITH URISYS 1100 WAS POSITIVE. THE SOFTWARE VERSION WAS 5.71. THE CUSTOMER USED A DIFFERENT VIAL OF TEST STRIPS AND THE RESULTS MATCHED THE PHYSICIAN'S RESULTS. THE CUSTOMER DID NOT PERFORM QC.