MalfunctionKDP

OHMEDA 1251 CONTINUOUS/INTERMITTENT VACUUM REGULATOR

Received Oct 21, 2019 · Event occurred Sep 5, 2019

Report 3007409280-2019-00044 · MDR key 9216422

Device

Generic name

Free Flow Vacuum Regulator

Model number

FROM1600

Product problems

  • Calibration Problem
  • Calibration Problem

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. INCOMING VISUAL INSPECTION DID NOT IDENTIFY ANY ANOMALIES. TECHNICAL VISUAL INSPECTION DID NOT IDENTIFY ANY ANOMALIES. DEVICE EVALUATION FOUND THE TIMING WAS OFF. THE TIMING WAS ADJUSTED. THE DEVICE WAS CALIBRATED. LEAK TEST AND SUCTION ACCURACY WAS TESTED AND PASSED. THE GAUGE TEST, LEAK TEST - SUPPLY/PATIENT SIDE, REGULATION TEST IN BOTH MODES ALL PASSED. THE FINAL VISUAL INSPECTION WAS GOOD. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED; HOWEVER, THE REPORTED ISSUE WAS MOST LIKELY DUE TO THE PRESSURE SENSOR FALLING OUT OF CALIBRATION. THIS WAS NOT RELATED TO THE PREVIOUS REPAIR. THIS TYPE OF EVENT WILL CONTINUE TO BE MONITORED.

Description of Event or Problem

REPORTEDLY, POST REPAIR, THE DEVICE GAUGE GOES ALL THE WAY AROUND WHEN THE SUCTION IS TURNED TO FULL AND ITS PORT IS OPEN. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT HARM REPORTED. NO ADDITIONAL INFORMATION AVAILABLE.