inforMED
MalfunctionDHX

ATTELLICA IM CARCINOEMBRYONIC ANTIGEN (CEA)

Received Oct 10, 2019 · Event occurred Sep 17, 2019

Report 1219913-2019-00204 · MDR key 9176997

Device

Generic name

Cea Immunoassay

Catalog number

10995523

Lot number

47916186

Product problems

  • No Apparent Adverse Event
  • No Apparent Adverse Event

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • Missing Value Reason
  • No Known Impact Or Consequence To Patient
  • Missing Value Reason

Narrative

Additional Manufacturer Narrative

SIEMENS FILED INITIAL MDR 1219913-2019-00204 ON OCTOBER 10, 2019. ADDITIONAL INFORMATION FROM 11/22/2019: CEA IS A DIRECT ASSAY AND AN INCORRECT HIGH RESULT WOULD BE DUE TO AN INCREASE IN RLU'S. SERVICE WAS PERFORMED FOR THE INSTRUMENT HOWEVER THE REAGENT PROBES 1 AND 2 WERE REPLACED WHICH WOULD NOT CORRECT HIGH RLU ON A DIRECT ASSAY. CEA DOES NOT USE REAGENT PROBE 2. POSSIBLE CAUSES OF GAINING RLU FOR A DIRECT ASSAY THE SPECIFICS CAN BE FOUND IN CB DOCS. ISSUES WITH WASH OR ASPIRATION, BASE DELIVERY (DRIPPING IN LUMINOMETER), SYSTEM CONTAMINATION - THIS ASSAY USES WATER FOR WASH/RINSE. REVIEW OF THE DATA DID NOT REVEAL A POTENTIAL PRODUCT PROBLEM. ON A FOLLOW UP VISIT THE CUSTOMER STATED THAT THEY ARE OPERATIONAL AND NO FURTHER INVESTIGATION IS REQUIRED. BASED ON THE INVESTIGATION, NO PRODUCT PROBLEM WAS IDENTIFIED. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Additional Manufacturer Narrative

THE CAUSE FOR THE DISCORDANT ATELLICA IM CEA RESULT IS UNKNOWN. SIEMENS HEALTHCARE DIAGNOSTICS IS INVESTIGATING. THE INSTRUCTION FOR USE (IFU) IN THE INTERPRETATION OF RESULTS SECTION STATES: "WARNING DO NOT USE THE ATELLICA IM CEA ASSAY AS A SCREENING TEST FOR DIAGNOSIS. NOTE DO NOT INTERPRET LEVELS OF CEA AS ABSOLUTE EVIDENCE OF THE PRESENCE OR THE ABSENCE OF MALIGNANT DISEASE. MEASUREMENTS OF CEA SHOULD ALWAYS BE USED IN CONJUNCTION WITH OTHER DIAGNOSTIC PROCEDURES, INCLUDING INFORMATION FROM THE PATIENT'S CLINICAL EVALUATION." THE INSTRUCTION FOR USE (IFU) IN THE EXPECTED VALUES SECTION STATES: "AS WITH ALL IN VITRO DIAGNOSTIC ASSAYS, EACH LABORATORY SHOULD DETERMINE ITS OWN REFERENCE INTERVAL FOR THE DIAGNOSTIC EVALUATION OF PATIENT RESULTS. 14 CONSIDER THESE VALUES AS GUIDANCE ONLY."

Description of Event or Problem

A CUSTOMER OBTAINED A DISCORDANT CEA RESULT ON THE ATELLICA IM 1300 ANALYZER. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN, WHO DID NOT QUESTION THE RESULT. THE CUSTOMER INDICATED THAT QUALITY CONTROL (QC) WAS FOUND TO BE OUTSIDE OF THE LABORATORY'S ACCEPTABLE RANGE ON 9/17/2019 ON ONE ATELLICA IM 1300 ANALYZER (IM00745). QC WAS REPEATED AND WHEN CONFIRMED TO BE IN RANGE CEA SAMPLES WERE REPEATED ON THE SAME INSTRUMENT. DISCORDANCE WAS FOUND BETWEEN THE INITIAL RESULT AND REPEAT RESULT ON ONE OF THESE SAMPLES. THE DISCORDANT SAMPLE WAS REPEATED AGAIN ON A DIFFERENT ATELLICA IM 1300 ANALYZER (IM00746) AND RESULT WAS SIMILAR TO THE FIRST REPEAT RESULT. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED AND THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT CEA RESULT.