INNER SHEATH WITH CERAMIC BEAK

Received Oct 4, 2019

Report 9610617-2019-00089 · MDR key 9156869

Device

Generic name

Inner Sheath With Ceramic Beak

Model number

27050CA

Catalog number

27050CA

Lot number

ZZ01

Product problems

  • Material Separation
  • Material Separation

Patient

Not reported

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS EVALUATED BY THE MANUFACTURER, KARL STORZ IN (B)(4). AS PER THE EVALUATION: THE CERAMIC TIP IS BROKEN IN ITSELF. HOWEVER, SLIGHT DISCOLORATION DUE TO FIRE DAMAGE IS VISIBLE AT THE DISTAL END. THE CERAMIC TIP HAS BEEN MECHANICALLY OVERLOADED - MOST LIKELY DROPPED WHAT PRE-DAMAGED THE CERAMIC TIP. AN INDICATOR FOR THIS IS THE CRACK FORMATION AT THE CERAMIC TIP. DAMAGES LIKE THIS HAPPEN IF THE CERAMIC INSERT IS HIT AGAINST HARD OBJECTS OR EDGES. DURING INTENDED USE, THERE IS NO POSSIBILITY TO APPLY FORCES TO ACHIEVE THE DEFECTS. THE WARNING NOTICES WITH THE CORRESPONDING PICTURES IN THE IFU INDICATE THIS.

Description of Event or Problem

AS PER A VIGILANCE REPORT FILED BY OUR PARENT COMPANY IN (B)(4), A PART OF THE TIP OF AN INNER SHEATH, ROTATING, WITH CERAMIC INSULATION, CAME OFF DURING INTERVENTION AND GOT LOST IN THE UTERUS. THE BROKEN PIECE WAS REMOVED FROM THE PATIENT.