COBAS B 101 HBA1C TEST
Received Oct 3, 2019 · Event occurred Sep 5, 2019
Report 1823260-2019-03562 · MDR key 9152603
Device
Generic name
Assay, Glycosylated Hemoglobin
Manufacturer
Roche DiagnosticsModel number
B 101 HBA1CCatalog number
08038694190
Lot number
929042-01
Product problems
- Non Reproducible Results
- Non Reproducible Results
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
RETENTION COBAS B101 HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 WERE INVESTIGATED. THE MANUFACTURER BATCH RECORD FOR COBAS B101 HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 SHOWED NO ABNORMALITIES. THE MANUFACTURER FINAL PRODUCT INSPECTION REPORT FOR COBAS B101 HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 SHOWED NO ABNORMALITIES AND ALL FEATURES PASSED. RETENTION DISCS COBAS B101 HBA1C DISC LOT 929042-01 AND LOT 930041-01 WERE MEASURED WITH BLOOD COMPARED TO REFERENCE METHOD: 4 REPLICATES 6 REPLICATES AT 2 LEVELS (WHOLE BLOOD LOW AND MIDDLE): 4 WHOLE BLOOD "LOW" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 929042-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. 4 WHOLE BLOOD "MIDDLE" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 929042-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. 6 WHOLE BLOOD "LOW" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 930041-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. 6 WHOLE BLOOD "MIDDLE" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 930041-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. THERE WERE NO DISCREPANT HBA1C RESULTS MEASURED WITH RETENTION MATERIAL HBA1C DISC LOT 929042-01 AND LOT 930041-01. DISCREPANT HBA1C RESULTS COULD NOT BE CONFIRMED. THE PROVIDED CUSTOMER DATA SHOWED DATA DEVIATION BETWEEN 1ST AND 2ND MEASUREMENT WAS 0.6-1.0% NGSP ON HBA1C LOT 929042-01. - RETENTION MATERIAL DATA SHOWED DATA DEVIATION BETWEEN 1ST AND 2ND MEASUREMENT WAS 0-0.1% NGSP ON HBA1C LOT 929042-01. - RETENTION MATERIAL TESTING: DATA DEVIATION BETWEEN HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 IS 0.1% NGSP. THE PROVIDED CUSTOMER'S DATA DEVIATION WAS 1.0%NGSP. - THE INVESTIGATION SHOWED THAT THE MEASUREMENT DATA OBTAINED WITH RETENTION MATERIAL HAS NO DEVIATION LIKE THE CUSTOMER COMPLAINED ABOUT. UPON REVIEW OF THE INSTRUMENT'S DATA LOGS, A WARNING MESSAGE OCCURRED AT THE SAME TIME THE COMPLAINED MEASUREMENTS WERE PERFORMED FOR THE FIRST SAMPLE. THE MESSAGE INDICATES ABNORMALLY HIGH HEMOGLOBIN. THIS MESSAGE OCCURS IF THE B101 INSTRUMENT DETECTS THAT THE HEMOGLOBIN CONCENTRATION IN THE TESTED BLOOD SAMPLE IS OVER THE MEASUREABLE RANGE (20G/DL). THE MESSAGE CAN ALSO OCCUR IF THE BLOOD SAMPLE ENTERED INTO THE DISC THROUGH THE INADEQUATE PORTION (HOLE). THE MESSAGE CAN ALSO OCCUR IF THE SAMPLE VOLUME IS TOO MUCH WHEN THE SAMPLE IS OVERFLOWED FROM THE CAPILLARY CHANNEL DUE TO THE IMPACT SUCH AS A DROP, THE HEMOGLOBIN CONCENTRATION IS OVER THE MEASURABLE RANGE. THE SAME WARNING MESSAGE OCCURRED MANY TIMES BETWEEN (B)(6) 2019 AND (B)(6) 2019.
Additional Manufacturer Narrative
THE REPORTER'S PRODUCT WAS REQUESTED FOR INVESTIGATION. THIS EVENT OCCURRED IN (B)(6).
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR THREE PATIENT SAMPLES TESTED WITH THE COBAS B 101 HBA1C TEST ON COBAS B 101 INSTRUMENT SERIAL NUMBER (B)(4). IT WAS ASKED, BUT IT IS NOT KNOWN IF ANY INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLES WERE REPEATED AND THE REPEAT MEASUREMENTS MATCHED THE CLINICAL CONDITION AND PREVIOUS RESULTS OF THE PATIENTS. THE REPORTER ALSO MENTIONED THAT THEY RECENTLY RECEIVE FREQUENT ERRORS ON THE COBAS B 101 INSTRUMENT. THE FIRST SAMPLE INITIALLY RESULTED WITH AN HBA1C VALUE OF 8.4 %, WHICH REPEATED AS 7.4 %. BOTH MEASUREMENTS WERE PERFORMED USING HBA1C TEST DISC LOT 929042-01. THE SECOND SAMPLE INITIALLY RESULTED WITH AN HBA1C VALUE OF 6.6 %, WHICH REPEATED AS 6.0 %. BOTH MEASUREMENTS WERE PERFORMED ON (B)(6) 2019 USING HBA1C TEST DISC LOT 929042-01. THE THIRD SAMPLE INITIALLY RESULTED WITH AN HBA1C VALUE OF 6.5 % WHEN MEASURED USING HBA1C TEST DISC LOT 929042-01 ON (B)(6) 2019. THE SAMPLE WAS REPEATED ON (B)(6) 2019 USING HBA1C TEST DISC LOT 930041-01, RESULTING WITH A VALUE OF 5.5 %.