Description of Event or Problem

TABLE TILTED WITH PATIENT ON IT AT THE HOSPITAL.

Additional Manufacturer Narrative

COLCOM SURGICAL DISPATCHED THEIR CERTIFIED SERVICE TECHNICIAN TO THE FACILITY TO MAKE A DETERMINATION OF THE COMPLAINT. THE TECHNICIAN FOUND A LEAK IN THE CHECK VALVE THAT CONTROLS THE TILT FUNCTION. HE ALSO FOUND A TORN O-RING IN THE MINI VALVE ASSEMBLY. HE REPLACED THE O-RING AND REPLACED THE CHECK VALVE. THE TABLE HAS BEEN PUT BACK IN SERVICE.

Remedial action

  • Repair