inforMED
InjuryILY

ANODYNE THERAPY

Received Sep 7, 2007 · Event occurred Jul 26, 2007

Report 1055581-2007-00014 · MDR key 908726

Device

Generic name

Infrared Lamp

Model number

480

Catalog number

498005

Lot number

NA

Product problems

  • Unknown (for use when the device problem is not known)

Patient

56 YR

  • Burn, Thermal
  • Treatment with medication(s)

Narrative

Additional Manufacturer Narrative

PT IS REPORTED TO HAVE DEVELOPED A BURN, APPROX 2" X 1 1/2" ON HER MEDIAL LOWER RIGHT CALF FOLLOWING TREATMENT WITH THE ANODYNE PROFESSIONAL SYSTEM. THE HEALTH CARE PROFESSIONAL ADMINISTERED TREATMENT FOR 22 MINUTES AT AN ENERGY SETTING OF 8. THIS IS WITHIN THE GUIDELINES PROVIDED IN THE IMPORTANT SAFETY INFO AND INSTRUCTIONS PROVIDED TO THE USER BY ANODYNE. THE UNIT INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION, AND DUTY CYCLE FAILURE WAS DETECTED. THE NUMBER OF INCIDENTS OF THIS TYPE REMAIN WELL WITHIN THE EXPECTED RATE FOR THIS PRODUCT BASED UPON THE NUMBER OF INCIDENTS REPORTED AND THE NUMBER OF UNITS IN DISTRIBUTION. COMPANY WILL CONTINUE TO MONITOR EVENTS OF THIS NATURE FOR TRENDING PURPOSE.

Description of Event or Problem

PT IS REPORTED TO HAVE DEVELOPED A BURN ON HER MEDIAL RIGHT CALF, FOLLOWING TREATMENT WITH THE ANODYNE THERAPY PROFESSIONAL SYSTEM. PT WAS BEING TREATED BY A HEALTH CARE PROFESSIONAL FOR PLANTAR FASCITIS. PT WAS TREATED BY HER DOCTOR WITH AN ANTIBIOTIC OINTMENT, AND HER WOUND IS HEALING.