CELL-DYN 22 TRI-LEVEL CONTROL
Received Sep 6, 2007 · Event occurred Jul 7, 2006
Report 2919069-2007-00143 · MDR key 907933
Device
Generic name
Cd22 Controls
Manufacturer
Abbott Diagnostic Div/celldynModel number
NACatalog number
93111-01
Lot number
0125
Product problems
- Unknown (for use when the device problem is not known)
Patient
NI YR
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
THE CUSTOMER REPORTED USING CELL-DYN 22 PLATELET CONTROL LOT 0125 SINCE 2006. THE CUSTOMER OPENED A NEW BOTTLE OF MEDIUM CONTROL THE FOLLOWING MONTH, AND OBSERVED AN INCREASE IN THE PLATELET PARAMETER. THE CUSTOMER WAS SENT NEW CELL-DYN 22 PLATELET CONTROL LOT 0126 AND THE PERFORMANCE WAS ACCEPTABLE. IT IS UNK IF THERE WAS IMPACT TO PT MANAGEMENT.
Additional Manufacturer Narrative
PLATELET RECOVERY VALUES WERE OUT OF RANGE HIGH DURING STABILITY TESTING FOR THE CELL-DYN 22 CALIBRATORS, LOTS 3099. 3098, AND 3100 AND CELL-DYN 22 CONTROL LOT 0125 FOR THE CD3200 AND/OR CD RUBY OPTICAL ANALYZERS. CELL-DYN 22 CALIBRATOR LOT 3098 DID NOT MEET SHELF LIFE STABILITY ON THE CELL-DYN 3200 ANALYZER AT ITS EXPIRY. CELL-DYN 22 CALIBRATOR LOT 3099 DID NOT MEET SHELF LIFE STABILITY AND THE 7 DAY OPEN VIAL STABILITY FOR THE PLATELET PARAMETER ON THE CELL-DYN RUBY ANALYZER. INTERNAL DATA INDICATED THAT CELL-DYN 22 CALIBRATOR LOT 3100 WAS EXPERIENCING THE SAME ISSUES AS IT APPROACHED THE 2006 EXPIRY DATE. CELL-DYN 22 CALIBRATORS AND CONTROLS ARE SUBJECTED TO STABILITY TESTING PER ABBOTT DEPARTMENT OPERATING PROCEDURES. THE CELL-DYN 22 CALIBRATORS AND CONTROLS CLAIM PRODUCT SHELF LIFE AND CONSECUTIVE DAY OPEN CONTAINER STABILITY. THE CALIBRATOR AND CONTROL VALUES MUST MEET THE REQUIREMENT WITHIN THE PUBLISHED ASSAY VALUES AND RANGES PER ABBOTT STABILITY TESTING PARAMETERS. THE VALUES RECOVERED DURING STABILITY TESTING ARE THE SAME AS PRODUCT RELEASE VALUES. AN INVESTIGATION WAS PERFORMED AND PROBABLE CAUSES WERE FOUND TO BE A COMBINATION OF THE FOLLOWING ISSUES: COMMON CAUSE PROCESS VARIABILITY OF THE RED BLOOD CELL FRAGMENTATION TO PLATELET LIKE PARTICLES AFFECTING MEAN PLATELET VOLUME, AND SPECIAL CAUSE VARIATION OF TIGHTENING THE PLATELET SPECIFICATION BY 50%. IT WAS DETERMINED THAT RED BLOOD CELL FRAGMENTATION IS ENHANCED BY MECHANICAL STRESS. THESE ISSUES DID NOT ADEQUATELY PREDICT AN INCREASE OF PLATELET COUNT THROUGH THE ESTABLISHED STABILITY CLAIMS. THE FORMULATION AND MANUFACTURING PROCESS USED IN THE MANUFACTURING AND TESTING OF THE CELL-DYN 22 CALIBRATOR AND CONTROLS ARE SPECIFIC TO BOTH THE CALIBRATORS AND CONTROLS AND DOES NOT IMPACT THE OTHER INSTRUMENT PLATFORMS, THEREFORE, THE IDENTIFIED PROBABLE CAUSE IS PRODUCT SPECIFIC. THE CORRECTIVE ACTION FOR THE CELL-DYN 22 CALIBRATORS AND CONTROLS IS TO RE-EVALUATE THE PLATELET SPECIFICATION, IMPROVE ASSAY ASSIGNMENT PROCEDURE FOR PLATELET PARAMETER, RE-VALIDATE THE PLATELET MANUFACTURING PROCESS TO INCREASE TO INCREASE THE MEAN PLATELET VOLUME AND CONDUCT MONTHLY VALUE ASSIGNMENT. THIS IS THE FINAL REPORT. END OF REPORT.
Remedial action
- Recall