inforMED
MalfunctionJSS

API® RAPID 20 E

Received Sep 17, 2019

Report 9615754-2019-00087 · MDR key 9076609

Device

Generic name

Api® Rapid 20 E Strip

Manufacturer

Biomerieux, Sa

Catalog number

20701

Lot number

1005589130

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Description of Event or Problem

AN INDUSTRY CUSTOMER FROM (B)(6) REPORTED A MISIDENTIFICATION OF SERRATIA MARCESCENS AS SERRATIA LIQUEFACIENS WHEN TESTING A QC STRAIN WITH API® RAPID 20 E 25STRIPS (REF. 20701) - LOT 1005589130. THE CUSTOMER STATED THEY TESTED A KNOWN SERRATIA MARCESCENS STRAIN AND RECEIVED AN IDENTIFICATION OF 95.5% S.LIQUEFACIENS (AND 3.9% S.MARCENSCENS). THE CUSTOMER ALSO USED SEQUENCING ON THIS STRAIN TO OBTAIN A PHYLOGENETIC TREE. THE SEQUENCING CONFIRMED THAT THE STRAIN WAS NOT S.LIQUEFACIENS. THE CUSTOMER THEN REPEATED THE RAPID 20 E TEST AND OBTAINED SERRATIA MARCESCENS. THERE WAS NO PATIENT INVOLVED AS THIS EVENT PERTAINS TO AN INDUSTRY QC. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

Additional Manufacturer Narrative

THIS REPORT WAS INITIALLY SUBMITTED FOLLOWING NOTIFICATION BY AN INDUSTRY CUSTOMER IN GERMANY OF A MISIDENTIFICATION OF SERRATIA MARCESCENS AS SERRATIA LIQUEFACIENS WHEN TESTING A QC STRAIN WITH API® RAPID 20 E 25STRIPS (REF. 20701) - LOT 1005589130. THE CUSTOMER STATED THEY TESTED A KNOWN SERRATIA MARCESCENS STRAIN AND RECEIVED AN IDENTIFICATION OF 95.5% S. LIQUEFACIENS (AND 3.9% S. MARCENSCENS). SEQUENCING METHODOLOGY CONFIRMED THE SPECIES SERRATIA MARCESCENS. BIOMÉRIEUX INTERNAL INVESTIGATION WAS CONDUCTED. REVIEW OF THE COMPLAINT IDENTIFIED: THE TESTED STRAIN WAS OBTAINED FROM A CONTAMINATION. THE CUSTOMER DOES NOT TEST THE RECOMMENDED STRAINS FOR QC. THE CUSTOMER PERFORMED RETEST WITH ANOTHER RAPID 20 E LOT NUMBER 1006885690. EXPIRATION DATE 22-APR-2020 AND OBTAINED A GOOD IDENTIFICATION TO SERRATIA MARCESCENS. COMPLAINTS TREND ANALYSIS IDENTIFIED ONE OTHER COMPLAINT ON RAPID 20 E REFERENCE 20701 FOR THE SERRATIA SPECIES. THIS COMPLAINT LED TO AN INVESTIGATION; HOWEVER, CUSTOMER PROBLEM WASN'T REPRODUCED INTERNALLY. NO OTHER COMPLAINT WAS REGISTERED ON THE LOT NUMBER 1005589130. REVIEW OF THE LOT RECORD INDICATED NO BATCH EVENT IMPACTING THE PERFORMANCE OF THE LOT 1005589130. AS API RAPID 20 E LOT NUMBER 1005589130 WAS EXPIRED AT THE TIME OF INVESTIGATION, NO FURTHER INVESTIGATION WAS POSSIBLE.