inforMED
MalfunctionIXL

DIAGNOST 76

Received Aug 31, 2007 · Event occurred Aug 1, 2007

Report 1217116-2007-00023 · MDR key 907119

Device

Generic name

Ixl - General Purpose R/f System

Model number

9870-694-12172

Catalog number

NA

Lot number

NA

Product problems

  • Break
  • Component Falling
  • Fail-Safe Design Failure

Patient

NA YR

  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED, A FOLLOW-UP REPORT WILL BE SENT TO THE FDA.

Description of Event or Problem

THIS REPORT IS BEING FILED UPON NOTIFICATION FROM OUR MFR IN ANOTHER COUNTRY OF AN EVENT THAT OCCURRED IN THAT COUNTRY IN ANOTHER CITY. DURING A BREAK BETWEEN PATIENT EXAMS WITH THE SYSTEM IN STANDBY AND IN ITS UPRIGHT POSITION (+90 DEGREES) THE TOW STEEL BANDS SUPPORTING THE SERIAL CHANGER / COUNTERWEIGHT BROKE. THE CARRIAGE FELL DOWN, OVERRAN THE MECHANICAL END STOP AND FINALLY STOPPED ON THE FLOOR. NO INJURY OCCURRED SINCE THERE WERE NO PERSONS PRESENT IN THE ROOM.