inforMED
MalfunctionFAR

ELECTRICOSURGICAL UNIT

Received Jul 29, 1994 · Event occurred Feb 23, 1994

Report 9064 · MDR key 9064

Device

Generic name

Esu, Bovie, Cartery

Model number

FORCE 4B

Patient

79 YR

Narrative

Description of Event or Problem

SUPERFICAL BURN ON LEFT HEEL. PATIENT LEG WAS RAISED AND ACCIDENTLY PLACED ON ESU PENCIL AND ACTIVATED.DEVICE NOT LABELED FOR SINGLE USE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.DEVICE SERVICED IN ACCORDANCE WITH SERVICE SCHEDULE. DATE LAST SERVICED: 01-JUL-93. SERVICE PROVIDED BY: USER FACILITY BIOMEDICAL/BIOENGINEERING DEPARTMENT. SERVICE RECORDS AVAILABLE.NO IMMINENT HAZARD TO PUBLIC HEALTH CLAIMED. DEVICE USED AS LABELED/INTENDED.DEVICE WAS EVALUATED AFTER THE EVENT. METHOD OF EVALUATION: ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED, ELECTRICAL TESTS PERFORMED, PERFORMANCE TESTS PERFORMED, VISUAL EXAMINATION. RESULTS OF EVALUATION: NONE OR UNKNOWN, NONE OR UNKNOWN, INCORRECT TECHNIQUE/PROCEDURE. CONCLUSION: USER ERROR CAUSED EVENT. CERTAINTY OF DEVICE AS CAUSE OF OR CONTRIBUTOR TO EVENT: YES. CORRECTIVE ACTIONS: USER EDUCATION PROVIDED. THE DEVICE WAS NOT DESTROYED/DISPOSED OF.