inforMED
MalfunctionCAN

FLOTEC FLOPAC REGULATOR

Received Oct 26, 2004 · Event occurred Jul 27, 2004

Report 1832475-2004-00004 · MDR key 904452

Device

Generic name

Pressure Regulator Intigrated With Cylinder Valve

Manufacturer

Flotec, Inc.

Model number

FP4-XX-B05-7005T2

Catalog number

FP4-XX-B05-7005T2

Lot number

M625000

Product problems

  • Decrease in Pressure

Patient

NO INFO

  • Low Oxygen Saturation

Narrative

Description of Event or Problem

DURING A PULM. REHAB SESSION PT ON STRESS TEST EXPERIENCED LOW SPO2 LEVELS WHEN OXYGEN CYLINDER RAN OUT OF OXYGEN. CONTENTS GAUGE SHOWED INACCURATE PRESSURE RESULTING IN CYLINDER TANK RUNNING OT OF OXYGEN BEFORE COMPLETION OF TEST.

Additional Manufacturer Narrative

GAUGE SHOWED AN INACCURATE PRESSURE AS PRESSURE IN CYLINDER TANK WAS REDUCED. CAUSE: DECAL RING WAS ASSEMBLED WITHOUT ADEQUATE DEBURN AND REQUIRED PRESET PROCESS DURING ASSEMBLY. PREVIOUSLY ISSUED CORRECTIVE ACTIONS HAS PREVENTED EVENTS SINCE THEY WERE PUT INTO PLACE. ADD'L STEPS: ECN ON USER INSTRUCTIONS AND TRANSFILLER INSTRUCTION CLARIFYING AND MODIFYING CHECKS ON FUNCTION OF GAUGE TO CHECK FOR DAMAGE AND/OR FUNCTION AND CALIBRATION.

Remedial action

  • Other
  • Inspection
  • Repair