TERUMO PERFUSION SYSTEM
Received Aug 24, 2007 · Event occurred Jul 25, 2007
Report 1828100-2007-00276 · MDR key 903246
Device
Generic name
Air Sensor
Manufacturer
Terumo Cardiovascular Systems Corp.Model number
5773Catalog number
5773
Product problems
- Sensor problems
Patient
YR
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE FAILURE WAS REPRODUCED BY MOVING/AGITATING ONE END OF THE AIR SENSOR CABLE. THE CABLE WAS FOUND TO HAVE AN INTERMITTENT CONNECTION ON ONE OF THE WIRES. DISASSEMBLY OF THE CABLE FOUND THE WIRE WAS LOOSE WITHIN THE SOLDER CUP AND AN INSUFFICIENT AMOUNT OF SOLDER WITHIN THE CUP. SOLDERING THE WIRE PROPERLY TO THE CUP RESULTED IN AN ELIMINATION OF THE PROBLEM.
Description of Event or Problem