inforMED
MalfunctionFRI

LIVONGO SCALE

Received Sep 11, 2019 · Event occurred Aug 8, 2019

Report 3011196194-2019-00015 · MDR key 8994548

Device

Generic name

Scale

Manufacturer

Livongo Health

Model number

BT004L

Catalog number

FG04087

Product problems

  • Temperature Problem
  • Temperature Problem

Patient

47 YR

  • Patient Problem/Medical Problem
  • Patient Problem/Medical Problem

Narrative

Additional Manufacturer Narrative

NO INJURY TO THE USER WAS REPORTED. THE SCALE WAS RETURNED TO THE MANUFACTURER, AND AN INVESTIGATION WAS PERFORMED. A PHYSICAL INSPECTION OF THE SCALE WAS PERFORMED AND FOUND A SMALL HOLE IN THE BACK OF THE SCALE. INITIAL FUNCTIONAL TESTING FOUND THAT THE SCALE DID NOT TURN ON. ROOT CAUSE WAS DETERMINED TO BE A USABILITY ISSUE, IN WHICH THE BATTERIES WERE INSERTED INTO THE SCALE INCORRECTLY, CAUSING A CAPACITOR TO SHORT CIRCUIT.

Description of Event or Problem

THE USER REPORTED THAT UPON BATTERY INSERTION THERE WAS A VISIBLE SPARK, AND A BURNING SMELL EMANATED FROM THE BATTERY COMPARTMENT. THE USER ALSO REPORTED THAT THE BATTERIES WERE WARM TO THE TOUCH AND A HOLE HAD MELTED IN THE BACK OF THE SCALE.