SUTURE HOOK CRESCENT MED LIMITED REUSE
Received Aug 16, 2007 · Event occurred Jul 11, 2007
Report 1017294-2007-00533 · MDR key 898683
Device
Generic name
Suture Hook
Manufacturer
Conmed LinvatecModel number
NACatalog number
C6370
Lot number
00872397
Product problems
- Tip breakage
- Device, removal of (non-implant)
Patient
YR
- Foreign body, removal of
Narrative
Additional Manufacturer Narrative
INVESTIGATION FINDINGS: THE DEVICE WAS RECEIVED FOR EVALUATION WITHOUT THE DETACHED PORTION. DURING A VISUAL EXAMINATION, THE NEEDLE WAS FOUND DETACHED AT THE WELD. FURTHER INSPECTION NOTED THAT THE OUTER TUBE WAS BENT. THE INFORMATION FOR USE (IFU) INFORMS THE USER OF THE FOLLOWING: CAUTIONS: AVOID LATERAL STRESSES TO THE INSTRUMENTS OR DEVICE FUNCTION MAY BE COMPROMISED AND UNINTENTIONAL PATIENT INJURY MAY RESULT. DO NOT USE IF PARTS ARE BROKEN, CRACKED OR WORN, OR DEVICE FUNCTION MAY BE COMPROMISED. AVOID MECHANICAL SHOCK OVER STRESSING THE INSTRUMENT WHICH MAY SHORTEN THE LIFE OF THE INSTRUMENT. INSTRUMENTS (HANDLE AND LIMITED REUSE HOOKS) SHOULD BE STORED IN THE SPECTRUM II STERILIZATION TRAY TO PROTECT THE FAILURES. AN INVESTIGATION REMAINS IN PROCESS FOR THIS FAILURE MODE. CONMED LINVATEC WILL CONTINUE TO MONITOR THIS PRODUCT FOR FAILURES.
Description of Event or Problem