CLINITEK STATUS+
Received Aug 30, 2019 · Event occurred Aug 7, 2019
Report 3002637618-2019-00098 · MDR key 8949859
Device
Product problems
- Electrical /Electronic Property Problem
- Electrical /Electronic Property Problem
Patient
Not reported
- No Patient Involvement
- No Patient Involvement
Narrative
Additional Manufacturer Narrative
INVESTIGATION COMPLETED: THE CLINITEK STATUS WAS RETURNED IN GOOD CONDITION WITH NO DAMAGE TO THE EXTERIOR OF THE INSTRUMENT. THE CORRECT POWER SUPPLY WAS RETURNED BUT NO BATTERIES WERE RETURNED. THE INSTRUMENT HAS D16 DAMAGE ON THE MAIN BOARD. D16 FAILURE CAUSED SYSTEM TO NOT OPERATE. THE D16 WAS THE SOURCE OF THE BURNING SMELL BUT IT IS NOT A FIRE RISK. D16 FAILS WHEN INCORRECT POWER IS CONNECTED TO THE DEVICE, EITHER THROUGH PLUGGING IN THE WRONG POWER SUPPLY (DEVICE USES A COMMON BARREL CONNECTOR) OR USING 3.5V RECHARGEABLE BATTERIES (DEVICE SPECIFIES USING COMMON 1.5V AA BATTERIES).
Additional Manufacturer Narrative
SIEMENS HAS REQUESTED THE CUSTOMER RETURN THE INSTRUMENT AND BATTERIES FOR INVESTIGATION. THE CAUSE OF THE EVENT IS UNKNOWN.
Description of Event or Problem
THE CUSTOMER REPORTED THAT THEIR CLINITEK STATUS+ WOULD NOT POWER ON. THEY TRIED A DIFFERENT POWER CORD AND OUTLET AND STILL NO POWER. WHEN THEY INSERTED BATTERIES, THE INSTRUMENT STARTED TO SMOKE FROM THE CASING OF THE INSTRUMENT, BEHIND THE DISPLAY. THE CUSTOMER STATED THERE WERE NO FLAMES OR INJURIES.
Remedial action
- Replace