inforMED
MalfunctionGDY

LAP SPONGE

Received Aug 29, 2019 · Event occurred Aug 1, 2019

Report 3010452421-2019-00001 · MDR key 8944176

Device

Generic name

Gauze/sponge, Internal, X-ray Detectable

Model number

1-418NS

Lot number

47481307

Product problems

  • Material Separation
  • Material Integrity Problem
  • Material Separation
  • Material Integrity Problem

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

AN INTERNAL COMPLAINT (B)(4) WAS RECEIVED INDICATING THAT AN X-RAY DETECTABLE ELEMENT OF A LAPAROTOMY SPONGE (PART 1-418NS, LOT 47481307) WAS EXPOSED AND CAME OUT WITH MINIMAL TENSION. THE DEFECTIVE SAMPLE WAS NOT AVAILABLE FOR RETURN. HOWEVER, THE REPORTING CUSTOMER PROVIDED A PHOTO THAT CONFIRMED THE REPORTED ISSUE. THE INVESTIGATION IS ONGOING AT THIS TIME. THIS REPORT WILL BE UPDATED WHEN NEW AND CRITICAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

THE X-RAY DETECTABLE PIECE IS EXPOSED ON THE SIDE OF THE SPONGE. THE LONGER STRIP CAME OUT WITHOUT MINIMAL TENSION.