MalfunctionBTM
AIRFLOW
Received Aug 27, 2019
Report 1314417-2019-00045 · MDR key 8936827
Device
Generic name
Manual Resuscitator
Manufacturer
Sunmed Holdings Llc.Model number
AF1140MB-TM5Lot number
318170
Product problems
- Incomplete or Missing Packaging
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE SUSPECT DEVICE WAS RETURNED, AND THE MASK PORT WAS CONFIRMED OUT OF ROUND VIA A VISUAL INSPECTION. THE MASK PORTS ARE 100% INSPECTED DURING ASSEMBLY THEREFORE THIS DEFECT OCCURRED DURING TRANSIT DUE TO PREVIOUS PACKAGING CONFIGURATIONS. THIS PRODUCT IS PRIOR TO A NEW PACKAGING CONFIGURATION WHICH USED A SLIGHTLY LARGER BOX AND NEW PACKAGING METHOD TO PREVENT MASK PORTS FROM DEFORMING.
Description of Event or Problem
THE CUSTOMER ALLEGES THAT "MASK CONNECTION IS OUT OF ROUND." NO OTHER DETAILS WERE PROVIDED AND NO PATIENT INJURY/HARM REPORTED.