Received Aug 7, 2007 · Event occurred Jul 26, 2007

Report 1825014-2007-00009 · MDR key 893496

Device

Generic name

Surgical Table

Model number

6500NB

Product problems

  • Mechanical Problem

Patient

Not reported

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

TABLE WENT INTO TRENDELENBERG WHEN ANOTHER FUNCTION BUTTON WAS PRESSED AT: MEDICAL CENTER.

Additional Manufacturer Narrative

THE SKYTRON 6500 TABLE WAS REPAIRED BY THE HOSPITAL'S BIOMED TECHNICIAN. HE FOUND THAT THE CAUSE OF THE MOVEMENT WAS DUE TO AN AGING SPOOL VALVE SPRING. HE REPLACED THE SPRING AND PLACED THE TABLE BACK IN SERVICE IN 2007.

Remedial action

  • Repair