IMAGE MANAGEMENT SYSTEM 660HD-E
Received Aug 23, 2019 · Event occurred Jul 27, 2019
Report 1643264-2019-00563 · MDR key 8927890
Device
Generic name
Device, Digital Image Storage, Radiological
Manufacturer
Smith & Nephew, Inc.Catalog number
72204668
Product problems
- No Display/Image
- Unexpected Shutdown
- No Display/Image
- Unexpected Shutdown
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE 660HD-E IMAGE MANAGEMENT SYSTEM HAD FOUND UNSATISFACTORY AS THE UNIT HAD AN ¿INTERMITTENT AUTO SHUTTING DOWN DURING SURGERY". 30 MINUTES DELAY AND A BACK UP WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO PATIENT INJURY WAS REPORTED.
Additional Manufacturer Narrative
THE DEVICE WAS RECEIVED FOR EVALUATION. THERE WAS NO RELATIONSHIP FOUND BETWEEN THE RETURNED DEVICE AND THE REPORTED INCIDENT. A VISUAL INSPECTION WAS PERFORMED ON THE PRODUCT AND NO ISSUE WAS OBSERVED. COMPLAINT OF SHUTTING DOWN BY ITSELF COULD NOT BE REPRODUCED. UNIT PASSED FUNCTIONAL TESTING DURING 6 HOUR BURN-IN INCLUDING CAPTURE AND PRINT FUNCTION. UNIT DID NOT SHUT DOWN DURING BURN-IN. ALL VIDEO INPUTS/OUTPUTS NORMAL AND UNIT'S DISPLAY PERFORMED AS EXPECTED. BOTH ANALOG AND DIGITAL VIDEO OUTPUTS WERE TESTED DURING BURN-IN. THE COMPLAINT WAS NOT CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED SINCE THE REPORTED MALFUNCTION COULD NOT BE DUPLICATED DURING THE PRODUCT EVALUATION PROCESS. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION OR WOULD NOT BE ABLE TO PERFORM AS INTENDED.