inforMED
MalfunctionGIT

DADE ACTIN FSL

Received Aug 6, 2007 · Event occurred Jul 20, 2007

Report 9610806-2007-00009 · MDR key 891514

Device

Generic name

Activated Ptt Reagent

Manufacturer

Dade Behring Gmbh

Model number

NA

Catalog number

B4219-2

Lot number

NA

Product problems

  • High Test Results

Patient

64 YR

  • Therapy/non-surgical treatment, additional

Narrative

Additional Manufacturer Narrative

THE CAUSE FOR THE ELEVATED APTT RESULT IS UNKNOWN. A DADE BEHRING REPRESENTATIVE SERVICED THE DEVICE, DID NOT DETECT AN INSTRUMENT ISSUE AND WAS UNABLE TO DUPLICATE THE CUSTOMER PROBLEM. THE INSTRUMENT IS PERFORMING AS EXPECTED.

Description of Event or Problem

A FALSELY ELEVATED APTT RESULT OF 70.2 SECONDS WAS OBTAINED ON A SAMPLE FROM A PATIENT ON HEPARIN. RESULTS WERE REPORTED TO THE PHYSICIAN BEFORE FOLLOWING LAB PROTOCOL TO REPEAT THE TEST. THE SAMPLE WAS RETESTED WITH RESULTS OF 49.8 AND 49.5 SECONDS. THE PATIENT'S HEPARIN DOSE HAD BEEN CHANGED BASED ON THE ORIGINAL RESULT OF 70.2 SECONDS, THEN THE DOSE WAS CHANGED AGAIN WHEN THE CORRECTED PTT RESULTS WERE REPORTED. NO PATIENT HARM WAS NOTED. THE CAUSE FOR THE ELEVATED APTT RESULT IS UNKNOWN.

Remedial action

  • Other