Received Aug 6, 2007 · Event occurred Jul 20, 2007
Report 9610806-2007-00009 · MDR key 891514
Device
Generic name
Activated Ptt Reagent
Manufacturer
Dade Behring GmbhModel number
NACatalog number
B4219-2
Lot number
NA
Product problems
- High Test Results
Patient
64 YR
- Therapy/non-surgical treatment, additional
Narrative
Additional Manufacturer Narrative
THE CAUSE FOR THE ELEVATED APTT RESULT IS UNKNOWN. A DADE BEHRING REPRESENTATIVE SERVICED THE DEVICE, DID NOT DETECT AN INSTRUMENT ISSUE AND WAS UNABLE TO DUPLICATE THE CUSTOMER PROBLEM. THE INSTRUMENT IS PERFORMING AS EXPECTED.
Description of Event or Problem
A FALSELY ELEVATED APTT RESULT OF 70.2 SECONDS WAS OBTAINED ON A SAMPLE FROM A PATIENT ON HEPARIN. RESULTS WERE REPORTED TO THE PHYSICIAN BEFORE FOLLOWING LAB PROTOCOL TO REPEAT THE TEST. THE SAMPLE WAS RETESTED WITH RESULTS OF 49.8 AND 49.5 SECONDS. THE PATIENT'S HEPARIN DOSE HAD BEEN CHANGED BASED ON THE ORIGINAL RESULT OF 70.2 SECONDS, THEN THE DOSE WAS CHANGED AGAIN WHEN THE CORRECTED PTT RESULTS WERE REPORTED. NO PATIENT HARM WAS NOTED. THE CAUSE FOR THE ELEVATED APTT RESULT IS UNKNOWN.
Remedial action
- Other