inforMED
MalfunctionLGC

IMMULITE 2000 XPI - HVG

Received Aug 20, 2019 · Event occurred Jul 23, 2019

Report 2432235-2019-00297 · MDR key 8913167

Device

Generic name

Immulite 2000 Xpi - Hvg

Catalog number

10381333

Lot number

303

Product problems

  • Low Test Results
  • Low Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE CAUSE FOR THE DISCORDANT FALSE NEGATIVE HVG RESULT IS UNKNOWN. SIEMENS IS INVESTIGATING THE ISSUE. MDR 2432235-2019-00296, MDR 2432235-2019-00298, AND MDR 2432235-2019-00299 WERE FILED FOR THE SAME EVENT.

Description of Event or Problem

ONE DISCORDANT, FALSE NEGATIVE RESULT WAS OBTAINED WITH HERPES SIMPLEX I/II IGG (HVG) LOT 303 ON IMMULITE 2000 XPI INSTRUMENT. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING THE SAME IMMULITE 2000 XPI INSTRUMENT AND HVG REAGENT LOT. THE REPEAT RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSE NEGATIVE HVG RESULT.

Additional Manufacturer Narrative

SIEMENS FILED INITIAL MDR ON 20-AUG-2019. ADDITIONAL INFORMATION (23-SEP-2019): SIEMENS HEADQUARTERS SUPPORT CENTER (HSC) REVIEWED THE DATA PROVIDED PREVIOUSLY BY THE CUSTOMER. HSC REVIEWED SAMPLE HANDLING BY THE CUSTOMER AND FOUND THE POSSIBILITY OF BUBBLES BEING PRESENT IN THE SAMPLE PRE ANALYSIS. REVIEW OF INTERNAL DATA WAS PERFORMED, AND NO ISSUES HAVE BEEN IDENTIFIED. PRE ANALYTICAL HANDLING CANNOT BE RULED OUT FOR THE NON-REPRODUCIBLE SAMPLE RESULTS. ADDITIONAL INFORMATION (16-OCT-2019): HSC REVIEWED SAMPLE ID: (B)(6). THE SAMPLE LEVEL SENSE WAS ANALYZED BUT FOUND WITHOUT ISSUES, THE DILUTION LEVEL SENSE WAS ALSO ANALYZED BUT NO ANOMALY WAS FOUND. ADDITIONALLY, THE REAGENT LEVEL SENSE DATA WAS ALSO REVIEWED, AND ERRORS WERE NOT FOUND. THE PRESENCE OF FOAM OR BUBBLES COULD NOT BE DETERMINED AS THE LEVEL SENSE ENCODER VALUES ARE CONSISTENT AND PERFORMING WITHOUT ISSUES. BASED ON THE INFORMATION PROVIDED, THE PROBABLE CAUSE IS INCONCLUSIVE. THE LEVEL SENSE PERFORMANCE OF THE INSTRUMENT IS WORKING AS EXPECTED. THERE IS NO EVIDENCE OF A PRODUCT NONCONFORMANCE. SAMPLE HANDLING OR PRE-ANALYTICAL CANNOT BE RULED OUT. THE CUSTOMER INDICATED THAT THEY HAD ANOTHER DISCORDANT HVG BUT DID NOT PROVIDE ANY DATA OR INFORMATION RELATED TO THE RESULTS. THE INSTRUMENT IS WORKING WITHOUT ISSUES. THE CAUSE FOR THE DISCORDANT FALSELY DEPRESSED HVG SAMPLE RESULT IS UNKNOWN. RELEVANT TESTS/LABORATORY DATA: UPDATED TO REFLECT A CORRECTION OF THE REPORTED SAMPLE ID. MDR 2432235-2019-00296-S1, MDR 2432235-2019-00298-S1, AND MDR 2432235-2019-00299-S1 WERE FILED FOR THE SAME EVENT.