Received Aug 19, 2019 · Event occurred Aug 8, 2019

Report 3013876692-2019-00014 · MDR key 8907221

Device

Generic name

Table And Attachments, Operating-room

Manufacturer

Holger Ullrich

Catalog number

100323C0

Product problems

  • Mechanical Problem
  • Use of Device Problem
  • Detachment of Device or Device Component
  • Insufficient Device Problem Information
  • Mechanical Problem
  • Use of Device Problem
  • Detachment of Device or Device Component
  • Insufficient Device Problem Information

Patient

Not reported

  • Laceration(s)
  • Injury
  • Laceration(s)
  • Injury

Narrative

Description of Event or Problem

MANUFACTURING REFERENCE #: (B)(4).

Additional Manufacturer Narrative

THE AFFECTED CLAMP WAS RETURNED TO THE FACTORY FOR INVESTIGATION. A FUNCTIONAL TEST WAS PERFORMED WITH THE AFFECTED CLAMP. IT WAS MOUNTED TO A MAQUET SIDE RAIL AND FIXED BY FASTENING THE DROP TOGGLE SCREW. THE FIRM SEAT WAS CHECKED AFTERWARDS AND FOUND TO BE ALRIGHT. NEXT AN ACCESSORY WAS ATTACHED TO THE RADIAL SETTING CLAMP AS DESCRIBED IN THE INSTRUCTIONS FOR USE (IFU). THE FIRM SEAT WAS CHECKED AND FOUND TO BE ALRIGHT, TOO. THE DIMENSIONS OF THE SIDE RAIL WHICH WAS USED WHEN THE INCIDENT OCCURRED WERE REQUESTED. THEY WERE PROVIDED, CHECKED AND FOUND TO BE WITHIN THE RANGE SPECIFIED IN THE IFU. THE SAME WAS DONE FOR THE WEIGHT OF THE USED ANESTHESIA FRAME. IT WAS CHECKED AND FOUND TO BE WITHIN THE RANGE SPECIFIED IN THE IFU. THE MALFUNCTION DESCRIBED BY THE USER COULD NOT BE REPRODUCED AT THE MANUFACTURER SITE. THEREFORE WE ASSUME A USE ERROR AS THE MOST PROBABLE ROOT CAUSE FOR THIS ISSUE. WE ASSUME THAT ONE OF THE SECURING ELEMENTS WAS NOT FASTENED PROPERLY. IN THE IFU IT IS DESCRIBED HOW THE CLAMP IS FIXED ON THE SIDE RAIL AND HOW ACCESSORIES ARE FIXED ON THE CLAMP. FURTHER THE USER IS WARNED AS FOLLOWS CONCERNING THE RISKS RELATED TO LOOSENED SECURING ELEMENTS: "..LOOSE OR LOOSENED SECURING ELEMENTS MAY CAUSE INJURIES. WHEN MOUNTING, AND AFTER EVERY ADJUSTMENT, TIGHTEN ALL OF THE LOCKING ELEMENTS (HANDLE SCREW, LOCKS, LEVERS, ETC.) OF THE PRODUCT. CHECK THE FIRM SEATING OF THE LOCKING ELEMENTS." IN THE COURSE OF THE INVESTIGATION TWO CRACKS WERE FOUND ON THE RADIAL SETTING CLAMP. IN THE PERFORMED TESTS, THESE CRACKS HAD NO IMPACT ON THE CORRECT FUNCTION OF THE PRODUCT. IN THE IFU THE USER IS TOLD TO PERFORM A VISUAL AND FUNCTIONAL INSPECTION PRIOR TO USE TO ENSURE CORRECT OPERATION. BESIDES SHE OR HE IS TOLD NOT TO USE A DEFECTIVE PRODUCT. MAQUET GMBH PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Additional Manufacturer Narrative

AS A RESULT OF AN INTERNAL REVIEW, IT WAS FOUND THAT THE IMPORTER AWARE DATE WAS INADVERTENTLY OMITTED.

Description of Event or Problem

MANUFACTURING REFERENCE #: (B)(4).

Description of Event or Problem

THE FOLLOWING WAS REPORTED. AN ANESTHESIA FRAME WAS MOUNTED TO AN OPERATING TABLE BY USING A RADIAL SETTING CLAMP. THE ANESTHESIA FRAME FELL ON THE PATIENT'S HEAD AND CAUSED A WOUND WHICH HAD TO BE STITCHED. MANUFACTURING REFERENCE # (B)(4).