inforMED
InjuryOCZ

RESCUE RAT TOOTH /ALLIGATOR GRASPING FORCEPS

Received Aug 16, 2019 · Event occurred Jul 20, 2019

Report MW5089136 · MDR key 8907146

Device

Generic name

Endoscopic Grasping / Cutting Instrument

Manufacturer

Endochoice, Inc.

Model number

DGC-383

Lot number

180915

Product problems

  • Entrapment of Device

Patient

22 YR

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Description of Event or Problem

AFTER SEVERAL PASSES DOWN THE SCOPE TO RETRIEVE PORT LODGED IN ESOPHAGUS, DEVICE COULDN'T BE DRAWN BACK THROUGH THE SCOPE CHANNEL. SCOPE WITH THE DEVICE IN THE CHANNEL HAD TO BE REMOVED FROM ESOPHAGUS AND THE END OF THE DEVICE CUT OFF TO ENABLE DEVICE TO BE REMOVED FROM SCOPE. ON INSPECTION, JAWS WERE TWISTED / OFFSET, WHICH WAS WHY TIP OF DEVICE WOULDN'T FIT BACK THROUGH SCOPE CHANNEL. FDA SAFETY REPORT ID# (B)(4).