inforMED
MalfunctionCEO

C8000 PHOSPHORUS

Received Jul 23, 2007 · Event occurred Dec 27, 2006

Report 2018433-2007-00042 · MDR key 883524

Device

Generic name

Quantitation Of Phosphorus In Human Serum, Plasma Or Urine

Model number

NA

Catalog number

7D71-20

Lot number

42020HW00

Product problems

  • Unknown (for use when the device problem is not known)

Patient

NI YR

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem

THE CUSTOMER STATED THAT PHOSPHATE CONTROL VALUES, RUN ON THE ARCHITECT C8000 ANALYZER, HAS SHIFTED UP AND IS OUT OF RANGE ON THE PHOSPHATE REAGENT LOT# 40460M200. THE CUSTOMER ADDED THE CORRECT CALIBRATOR VALUE IN THE ASSAY FILE AND RECEIVED ACCEPTABLE RESULTS. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED.

Remedial action

  • Other