MalfunctionBSZ
ADU
Received Jul 19, 2019
Report 9610105-2019-00005 · MDR key 8807603
Device
Product problems
- Mechanical Problem
- Gas/Air Leak
- Mechanical Problem
- Gas/Air Leak
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
FOR THE REPORTED EVENT, A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED EVENTS. TO RESOLVE THE REPORTED ISSUE, THE COMPACT BLOCK WAS REPLACED.
Description of Event or Problem
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT. A REVIEW OF THE EVENT INDICATED THAT MODEL ADU-98 ANESTHESIA GAS MACHINE EXPERIENCED A MALFUNCTION RESULTING IN A LEAK CAUSING A COMPLETE LOSS OF ALL VENTILATION MODES. THE REPORTED EVENT DID NOT INVOLVE A PATIENT.