inforMED
MalfunctionILZ

DELUXE TRAPEZE

Received Jul 13, 2007 · Event occurred Jan 31, 2007

Report 2921578-2007-00002 · MDR key 880637

Device

Generic name

Traction

Model number

3071

Catalog number

3071

Product problems

  • Detachment Of Device Component
  • Component Falling

Patient

50 YR

  • Abrasion
  • Swelling

Narrative

Additional Manufacturer Narrative

THE PRODUCT IS DESIGNED TO WITHSTAND A 250 LB PATIENT, THE S-HOOK THAT FAILED ON THE TRAPEZE (MASTER LINK TO CONNECT CHAIN TO TRAPEZE AND TO CLAMP) IS DESIGNED TO WITHSTAND 300 LBS. IT APPEARS TO BE A USER ERROR, POSSIBLY DUE TO REVIEW OF DEVICE BEFORE AND AFTER INSTALLATION ON HOSPITAL BEDS.

Description of Event or Problem

AS PER THE HOSPITAL REPRESENTATIVE, THE PATIENT WAS LYING IN BED USING THE TRAPEZE TO SCOOT UP ON THE HOSPITAL BED. THE TRAPEZE SUB-COMPONENT "D" RING FAILED AND PART FELL ONTO PATIENT'S HEAD. PATIENT EXPERIENCED SWELLING AND ENDURED A SMALL ABRASION ON TOP OF HIS HEAD.