MalfunctionDWH
POWERPRO BATTERY RECIPROCATING SAW
Received Jul 10, 2007 · Event occurred Jun 28, 2007
Report 1017294-2007-00460 · MDR key 879835
Device
Generic name
Powerpro Battery System
Manufacturer
Conmed LinvatecModel number
NACatalog number
PRO5400
Lot number
NA
Product problems
- Device Inoperable
Patient
YR
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
INVESTIGATION FINDINGS: THE HANDPIECE WAS RECEIVED FOR EVALUATION AND DURING TESTING WAS FOUND TO HAVE THE POTENTIAL TO OPERATE ON ITS OWN RELATED TO CONTROLLER FAILURE. AN INVESTIGATION FOR THIS FAILURE MODE REMAINS IN PROCESS. CONMED LINVATEC WILL CONTINUE TO MONITOR THIS PRODUCT FOR FAILURES. THERE ARE NO REPORTED INJURIES RELATED TO THIS FAILURE MODE.
Description of Event or Problem
THIS HANDPIECE WAS RETURNED WITH THE REPORT THAT IT WOULD NOT OPERATE. THERE ARE NO INJURIES OR DELAYS RELATED TO THIS EVENT.