inforMED
MalfunctionJIL

CHEMSTRIP ® 10 MD

Received Jul 11, 2019 · Event occurred Jun 11, 2019

Report 1823260-2019-02534 · MDR key 8780585

Device

Generic name

Urine Test Strips

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

03260763160

Lot number

33828902

Product problems

  • High Test Results
  • High Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE CUSTOMER DOES NOT RUN QC. THE CUSTOMER USED 3 VIALS OF TEST STRIPS WITH THE SAME LOT NUMBER. TWO VIALS HAVE BEEN DISCARDED. THE ONE REMAINING VIAL WAS REQUESTED FOR INVESTIGATION.

Description of Event or Problem

THE INITIAL REPORTER COMPLAINED OF FALSE POSITIVE NITRITE RESULTS FOR 5 PATIENTS TESTED WITH CHEMSTRIP 10 MD URINE STRIPS ON URISYS 1100 ANALYZER SERIAL NUMBER (B)(4) ON (B)(6) 2019 AND (B)(6) 2019. WHEN THE METER RESULT WAS "POSITIVE" FOR ALL 5 PATIENTS, ALL 5 PATIENTS WERE GIVEN ANTIBIOTICS. NO SPECIFIC RESULTS WERE PROVIDED. A VISUAL READING WAS NOT PERFORMED. THE SAMPLES WERE SENT TO THE LABORATORY AND CULTURED. TWO DAYS AFTER THE METER RESULTS, THE CULTURE RETURNED AS NEGATIVE FOR NITRITES FOR ALL 5 PATIENTS. WHEN THE NEGATIVE RESULTS WERE RETURNED, THE ANTIBIOTICS WERE STOPPED. THE PHYSICIAN STATED NONE OF THE PATIENTS WERE ADVERSELY AFFECTED DUE TO THE ANTIBIOTICS. NO ADVERSE EVENT OCCURRED.

Additional Manufacturer Narrative

INVESTIGATION DID RETENTION TESTING ON ANOTHER LOT 33828900 AND ON THE CUSTOMER LOT. NO FALSE POSITIVE RESULTS WERE OBSERVED WITH METER OR STRIPS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.