RAPIDLYTE 3 ML L/S SYRINGE FILTER CAP
Received Jul 5, 2019 · Event occurred Jun 11, 2019
Report 3002637618-2019-00077 · MDR key 8764885
Device
Generic name
Rapidlyte 3 Ml L/s Syringe Filter Cap
Manufacturer
Siemens Healthcare DiagnosticsCatalog number
10707031
Product problems
- Material Puncture/Hole
- Use of Device Problem
- Material Puncture/Hole
- Use of Device Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
INVESTIGATION COMPLETED: REVIEW OF MANUFACTURING RECORDS FOR LOT # 3673346 REVEALED NO DISCREPANCIES OR ANOMALIES. A TOTAL OF THREE USED SYRINGES WERE RECEIVED FOR ANALYSIS. NO ORIGINAL PACKAGING WAS RETURNED. VISUAL EXAMINATION OF THE SYRINGES REVEALED HOLES IN THE MOLDING GATES OF TWO SYRINGE BARRELS. A SIMILAR HOLE WAS FOUND IN THE SHOULDER OF THE THIRD SYRINGE'S BARREL. IN RESPONSE TO THESE OBSERVATIONS, NOTIFICATIONS WERE PROVIDED TO THE SUPPLIER FOR THE SYRINGE BARREL COMPONENT AS WELL AS THE MANUFACTURER'S PERSONNEL. COMPLAINT DATA WILL CONTINUE TO BE MONITORED FOR NEW INFORMATION, WITH FURTHER ACTIONS TAKEN ACCORDINGLY.
Additional Manufacturer Narrative
(B)(6) 2019: UPDATE: BIOHAZARD PACKAGING IS REQUIRED FOR THE SYRINGES TO BE SENT TO THE SUPPLIER FOR INVESTIGATION. THE CONTAINERS THAT WERE SHIPPED WERE TOO SMALL FOR THE USED SYRINGES. SIEMENS IS WAITING FOR THE LARGER CONTAINERS TO BE SENT.
Additional Manufacturer Narrative
SIEMENS HAS REQUESTED THE SYRINGES BE PROVIDED FOR INVESTIGATION. THE CUSTOMER STATED THAT THE OPERATOR WASHED THEIR EYES WITH EYE WASH SOLUTION AND DID NOT SEEK MEDICAL INTERVENTION.
Description of Event or Problem
THE CUSTOMER REPORTED THAT A RAPIDLYTE 3 ML L/S SYRINGE SPRAYED BLOOD WHILE ASPIRATING A SAMPLE. THE BLOOD SPRAYED IN THE OPERATOR'S EYE. THE OPERATOR WAS NOT WEARING SAFETY GLASSES.
Additional Manufacturer Narrative
(B)(6) 2019: UPDATE: THE SUPPLIER HAS INDICATED THERE WAS A CLEARANCE DELAY THAT DELAYED THE INVESTIGATION. THE SHIPMENT HAS NOW BEEN RELEASED AND IS ON ITS WAY TO THE INVESTIGATION SITE.