ARTHREX
Received Jul 2, 2019 · Event occurred Jun 14, 2019
Report 3032391-2019-00014 · MDR key 8754915
Device
Generic name
Scorpion¿ Needle, For Use With Scorpion¿ Suture Passer, (blue)
Manufacturer
Medline RenewalCatalog number
AR13990NR
Lot number
414465
Product problems
- Break
- Entrapment of Device
Patient
60 YR
- Foreign Body In Patient
- No Known Impact Or Consequence To Patient
- Foreign Body In Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
IT WAS REPORTED THAT THE TIP OF THE REPROCESSED ARTHREX SCORPION¿ SUTURE PASSER (BLUE) BROKE OFF INTO THE PATIENT DURING AN UNIDENTIFIED LEFT SHOULDER SURGERY. REPORTEDLY, THE INCIDENT HAPPENED WHILE TRYING TO PASS A SUTURE THROUGH THE SHOULDER. IT WAS ADDED THAT THE SURGEON COULD NOT FIND THE TIP AND ASSUMED IT WAS LEFT INSIDE THE PATIENT. THE PATIENT WAS NOTIFIED POST-OP AND REPORTEDLY REFUSED AN X-RAY TO DETERMINE IF THE TIP WAS RETAINED IN THE SHOULDER. DUE TO THE REPORTED EVENT, THIS MEDWATCH IS BEING FILED. THE SAMPLE IS NOT AVAILABLE TO BE RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD FOR THE REPORTED LOT NUMBER WAS PERFORMED AND THIS INDICATED THAT ALL PROCESSES WERE CONDUCTED AS REQUIRED AT THE TIME THE LOT WAS PROCESSED. A ROOT CAUSE FOR THE REPORTED ISSUE COULD NOT BE DETERMINED AT THIS TIME. NO ADDITIONAL INFORMATION IS AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED THAT THE TIP OF THE REPROCESSED ARTHREX SCORPION¿ NEEDLE PASSER BROKE OFF INTO THE PATIENT AND COULD NOT BE FOUND.