MERITS HEALTH PRODUCTS
Received Jul 2, 2019 · Event occurred May 17, 2019
Report 1000221992-2019-00001 · MDR key 8752687
Device
Product problems
- Mechanical Problem
- Mechanical Problem
Patient
65 YR
- Injury
- Injury
Narrative
Additional Manufacturer Narrative
WE RECEIVED THE PROBLEM DESCRIPTION OF MEDWATCH REPORT AND RELATED PHOTOS FROM THE IMPORTER. THE DEVICE WAS YET NOT RETURNED BACK. BASE ON THE PROVIDED SERIAL NUMBER, THE FINAL INSPECTION REPORT OF THIS SPECIFIC DEVICE WAS REVIEWED AND THE DEVICE DID COMPLY WITH OUR QUALITY CRITERIA. BASED ON THE PROVIDED PHOTOS, THE POSSIBLE REASON OF THIS EVENT MIGHT BE THAT THE FOOTPLATE LENGTH WAS SET AT THE LOWEST POSITION AND THE USER DID NOT RAISE UP THE POWER FOOTPLATE TO ENSURE ENOUGH GROUND CLEARANCE WHEN RIDING FROM THE SIDEWALK TO THE PAVEMENT. BASE ON THE PROVIDED INFORMATION, THE USER DID NOT USE THE SAFETY BELT AT THAT TIME. THIS HAS CAUSED HIM FELL OUT THE CHAIR WHEN THIS EVENT HAPPENED. IN ORDER TO ENSURE THAT THE USER IS VERY CLEARLY AWARE THAT THEY NEED TO PAY ATTENTION TO ENSURE ENOUGH GROUND CLEARANCE, WE WILL UPDATE OUR OWNER'S MANUAL AND ADD ADDITIONAL WARNING LABEL.(SEE APPENDIX IV) IF THE DEVICE IS RETURNED FOR FURTHER INVESTIGATION, IN CASE THERE IS ANY ADDITION ACTION REQUIRED, WE WILL SUBMIT FOLLOW UP REPORT. (B)(4).
Description of Event or Problem
THE PATIENT CLAIMED THE FOOTPLATE DROPPED DOWN AND CAUGHT ON THE CONCRETE WHICH CAUSED HIM TO FLIP OUT OF HIS CHAIR SUSTAINING INJURIES. THE PATIENT WAS APPARENTLY NOT WEARING HIS SEAT BELT AND UPON EXAMINATION OF THE FOOTPLATE IT HAD BEEN ADJUSTED TO LENGTHEN IT WHICH LESSENED THE GROUND CLEARANCE. THE DEALER REPORTED THAT HE WENT TO THE DOCTOR THE NEXT DAY AND THEY JUST GAVE HIM INSTRUCTIONS ON HOW TO ATTEND TO HIS SCRAPES. PATIENT NOW CLAIMS THAT HE HAS HEADACHES AND NEEDS PLASTIC SURGERY FOR HIS NOSE. HE IS SEEKING COMPENSATION. THE DEALER IS UNWILLING TO SEND THE MEDICAL RECORDS TO US DUE TO HIPPA REGULATIONS. FOOTPLATE WAS REPLACED BY DEALER AND PATIENT IS CURRENTLY USING CHAIR.