INNER TUBE WITH CERAMIC BEAK
Received Jun 28, 2019 · Event occurred Apr 16, 2019
Report 9610617-2019-00053 · MDR key 8745298
Device
Generic name
Inner Tube With Ceramic Beak
Manufacturer
Karl Storz Se & Co. KgModel number
27050XBCatalog number
27050XB
Lot number
NS01
Product problems
- Material Separation
- Material Separation
Patient
78 YR
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
WE EVALUATED THE INNER TUBE AND FOUND A PORTION OF THE CERAMIC BEAK IS BROKEN OFF, THERE ARE TOOL MARKS ON THE DISTAL END OF METAL SHAFT NEAR THE BEAK, AND THE SHAFT IS DENTED. IN ADDITION, WE FOUND THAT THIS INNER TUBE HAS BEEN 3RD PARTY REPAIRED; THE CERAMIC BEAK IS WHITE WHEN IT SHOULD BE GRAY IN COLOR, AND THE CERAMIC MATERIAL IS THINNER THAN OUR GRAY CERAMIC MATERIAL. THE MOST LIKELY CAUSE OF THIS TYPE OF BREAKAGE OF THE CERAMIC BEAK IS MECHANICAL FORCE.
Description of Event or Problem
ALLEGEDLY, A PORTION OF THE INNER TUBE'S CERAMIC BEAK BROKE DURING A TURBT PROCEDURE; THE BROKEN PIECE WAS SUBSEQUENTLY FOUND IN A SPECIMEN CUP BY THE LAB.