inforMED
MalfunctionMKZ

AMPLICOR CT/NG TEST

Received Sep 28, 2006

Report 2243471-2006-00011 · MDR key 873321

Device

Generic name

In Vitro Diagnostic Nucleic Acid Test

Model number

NA

Catalog number

20759392018

Lot number

H02039

Product problems

  • Device Difficult to Program or Calibrate

Patient

*

  • Unknown (for use when the patient's condition is not known)

Narrative

Additional Manufacturer Narrative

THE PACKAGE INSERT THAT ACCOMPANIES THE AMPLICOR CT/NG TEST STATES IN THE INSTRUCTIONS FOR USE SECTION, PART D AMPLIFICATION, STEP 3 "DURING RUN SET-UP, SET THE REACTION VOLUME TO 100UL BY FIRST CHANGING THE RAMP SPEED FROM THE 9600 MODE TO THE MAX MODE. TO DO THIS, CURSOR DOWN TO RAMP SPEED AND SELECT MAX". APPLIED BIOSYSTEMS GENEAMP PCR SYSTEM 9700 THERMAL CYCLERS WITH THE ALUMINUM HEAT BLOCK WILL NOT ACCEPT THIS SETTING. ONLY THERMAL CYCLERS EQUIPPED WITH THE GOLD-PLATED HEAT BLOCK WILL ACCEPT THE RAMP SPEED SETTING OF MAX. THE PURPOSE OF PROVIDING THE PRODUCT BULLETIN WAS TO PROVIDE ADDITIONAL INFORMATION ON THIS SETTING REQUIREMENT. ADD'L LOT # G14950.

Description of Event or Problem

IN JUNE 2006, ROCHE MOLECULAR SYSTEMS, INC. PROVIDED A PRODUCT BULLETIN CLARIFYING THE USE OF THE GOLD PLATED HEAT BLOCK AND THE ALUMINUM HEAT BLOCK ON THE APPLIED BIOSYSTEMS GENEAMP PCR SYSTEM 9700 THERMAL CYCLER. IN RESPONSE TO THE THE BULLETIN, THE CUSTOMER REPORTED THEY HAD BEEN USING THE ALUMINUM HEAT BLOCK WHEN PERFORMING THE AMPLICOR CT/NG TEST. THE ALUMINUM HEAT BLOCK WILL NOT ACCEPT A MAXIMUM RAMP SPEED WHICH IS REQUIRED TO PERFORM THE AMPLICOR CT/NG TEST. THE CURRENT AMPLICOR CT/NG TEST PACKAGE INSERT DESCRIBES THE SETTING OF THE TEMPERATURE RAMP SPEED TO MAXIMUM BUT DOES NOT IDENTIFY THE HEAT BLOCK TO BE USED.