COMBUR-10-M 100 STR
Received Jun 19, 2019 · Event occurred May 29, 2019
Report 1823260-2019-02243 · MDR key 8711688
Device
Generic name
Urine Test Strips
Manufacturer
Roche DiagnosticsModel number
COMBUR-10-MCatalog number
ASKU
Lot number
37378004
Product problems
- High Test Results
- High Test Results
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
EXPIRATION DATE WAS UPDATED. THE RETENTION MATERIAL, LOT 37378000, WAS MEASURED ON INVESTIGATION UNIT URISYS 1800 AND U 411 ANALYZER WITH 0-NATIVE URINE AND A NITRITE-DILUTION-SERIES AND WAS VISUALLY CHECKED. THE RESULTS OF THE MEASUREMENTS FULFILL THE REQUIREMENTS. NO FALSE POSITIVE RESULTS WERE OBSERVED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
AFTER CALIBRATION, THERE WERE NO MORE DISCREPANCIES WITH NITRITE RESULTS. THIS EVENT OCCURRED IN (B)(6).
Description of Event or Problem
THE INITIAL REPORTER STATED THAT THEY HAVE RECEIVED DISCREPANT HIGH RESULTS FOR AN UNSPECIFIED NUMBER PATIENT SAMPLES TESTED FOR NITRITES ON A COBAS U 411 ANALYZER. NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER STATED THAT THEY TESTED 142 URINE SAMPLES PER DAY OVER A 6 DAY PERIOD USING THE SAME TEST STRIP LOT NUMBER. WHEN TESTING THE SAMPLES ON THE U 411 ANALYZER, THE NITRITE RESULTS WILL BE POSITIVE. WHEN RUNNING TEST STRIPS MANUALLY ON THESE SAMPLES, THE VISUAL READING IS NEGATIVE FOR NITRITE. NO ADVERSE EVENTS WERE ALLEGED TO HAVE OCCURRED WITH THE PATIENTS. THE SERIAL NUMBER OF THE U 411 ANALYZER IS (B)(4). THE REPORTER'S PRODUCT WAS REQUESTED FOR INVESTIGATION.