inforMED
MalfunctionLGH

ION NITRIDED BLADE

Received Jun 11, 2019 · Event occurred May 14, 2019

Report 0001811755-2019-01950 · MDR key 8687208

Device

Generic name

Instrument, Cast Removal, Ac-powered

Catalog number

0940023000

Lot number

UNKNOWN

Product problems

  • Appropriate Device Problem Term/Code Not Available
  • Appropriate Device Problem Term/Code Not Available

Patient

Not reported

  • Abrasion
  • Abrasion

Narrative

Additional Manufacturer Narrative

INVESTIGATION RESULTS INDICATE THAT EXCESSIVE USE OF THE BLADE ALONG WITH USING THE HANDPIECE IN A CONTINUOUS LINEAR MOTION MAY HAVE CONTRIBUTED TO THE ABRASION. THE IFU FOR CAST CUTTER BLADES HAS THE FOLLOWING WARNING; ¿CUT WITH AN UP AND DOWN MOTION. STRAIGHT CUTTING MAY CUT OR BURN THE PATIENT.¿ THE QUALITY INVESTIGATION IS COMPLETE.

Description of Event or Problem

IT WAS REPORTED THAT FOLLOWING THE REMOVAL OF A CAST IT WAS OBSERVED THAT THE CUSTOMER WAS CUT OR BURNED. NO FURTHER INFORMATION WAS PROVIDED.

Additional Manufacturer Narrative

A FOLLOW UP REPORT WILL BE FILED ONCE THE QUALITY INVESTIGATION IS COMPLETE.

Description of Event or Problem

IT WAS REPORTED THAT FOLLOWING THE REMOVAL OF A CAST IT WAS OBSERVED THAT THE CUSTOMER WAS CUT OR BURNED. NO FURTHER INFORMATION WAS PROVIDED.