inforMED
InjuryKSS

GPS MINI KIT WITH ACD-A & BD -A/BLOOD D

Received Jun 3, 2019 · Event occurred May 8, 2019

Report 0001825034-2019-02404 · MDR key 8662532

Device

Generic name

Supplies, Blood-bank

Catalog number

800-0670A

Lot number

UNK

Product problems

  • Separation Problem
  • Separation Problem

Patient

Not reported

  • No Information
  • No Information

Narrative

Additional Manufacturer Narrative

(B)(4). FOREIGN - (B)(6). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT THE DEVICE DID NOT REMOVE ENOUGH PLASMA.

Additional Manufacturer Narrative

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL EXAMINATION OF THE RETURNED PRODUCT IDENTIFIED THERE IS DRIED BLOOD IN THE CANISTER. THERE ARE NO BLOCKAGES IN THE SILICONE COILED TUBE. THE PORTS ARE NOT DAMAGED OR CRACKED. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. A SUMMARY OF THE INVESTIGATION HAS BEEN SENT TO THE COMPLAINANT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.