inforMED
InjuryIZZ

COMPAX 40E

Received Jun 14, 2007 · Event occurred Jun 2, 2007

Report 2126677-2007-00017 · MDR key 864970

Device

Model number

2208203

Catalog number

NA

Lot number

NA

Product problems

Patient

73 YR

  • Fall
  • Bone Fracture(s)

Narrative

Additional Manufacturer Narrative

A GE FIELD ENGINEER THAT WAS DISPATCHED TO THE SITE NOTICED THAT THE VINYL COVER ON THE FOOT PEDAL FOR THE TABLE WAS LOOSE. IT WAS SURMIZED THAT THIS VINYL COVER JAMMED THE FOOT PEDAL IN THE "RELEASE" POSITION NOT ALLOWING THE TABLE LOCKS TO ENGAGE. THE FIELD ENGINEER RE-GLUED THE VINYL COVER TO THE FOOT PEDAL AT THE SITE.

Description of Event or Problem