RELIANCE MEDICAL PRODUCTS, INC
Received May 13, 2019 · Event occurred Dec 31, 2018
Report 1523545-2019-00001 · MDR key 8603307
Device
Generic name
Model 6200 Examination And Treatment Chair
Manufacturer
Reliance Medical Products, Inc.Model number
6200LProduct problems
- Mechanical Problem
- Mechanical Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
RELIANCE MEDICAL RECEIVED A COMPLAINT ALLEGING A #18 HEADREST FAILED ON A 6200L MODEL CHAIR. NO SERIES INJURY TOOK PLACE. (B)(6) MEDICAL CENTER BIOMED STATED THAT WHEN PRESSURE IS APPLIED THE HEADREST IT MAY COLLAPSE BACKWARDS. RELIANCE TECHNICAL SERVICE FOLLOWED UP AND IN-SERVICED THE CUSTOMER ON PROPER INSTRUCTIONS TO ENSURE THE HEADREST IS LOCKED. RELIANCE TECHNICIAN EXPLAINED THE NO. 18 HEADREST IS DESIGNED SO IT CAN BE OPERATED WITH JUST ONE HAND. TO UNLOCK, SQUEEZE THE CLEAR PLASTIC ACTUATOR INTO LOCK HOUSING. THIS WILL ALLOW BOTH PIVOTS OF THE ARTICULATING ARM TO ROTATE SO THE HEADREST CAN BE MOVED TO A NEW DESIRED POSITION. ONCE IN THE NEW POSITION, RELEASE THE ACTUATOR. MAKE SURE THE ACTUATOR EXPANDS TO ITS ORIGINAL POSITION FULLY EXPOSING BLACK DOT ON HANDLE TO INSURE A POSITIVE LOCK. THE HEADREST PAD IS FREE TO ROTATE AT ALL TIMES. THE CUSTOMER DISPOSED OF THE HEADREST AND WAS UNABLE TO PROVIDE THE HEADREST BACK TO RELIANCE FOR FURTHER INVESTIGATION. RELIANCE IS UNABLE TO CONFIRM THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AND REPORTING ON AN ALLEGED MALFUNCTION. THIS COMPLAINT WILL BE LOGGED AS COMPLAINT #(B)(4). THE CUSTOMER HAS REPLACED THE 6200L MODEL CHAIR WITH FX920L. THE FX920L FUNCTIONS AS DESIGNED.
Description of Event or Problem
RELIANCE MEDICAL RECEIVED A COMPLAINT ALLEGING A #18 HEADREST FAILED AND A PATIENT EXPERIENCED WHIPLASH. RELIANCE FOLLOWED UP CONCERNING THE ALLEGATION OF WHIPLASH AND THE CUSTOMER DID NOT PROVIDE THE NECESSARY INFORMATION TO DETERMINE THE SEVERITY OF HARM OF THE ALLEGED INJURY. RELIANCE DOES NOT CONSIDER THIS A SERIOUS INJURY. IF ANY NEW ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED, THE INJURY WILL BE CHARACTERIZED ACCORDINGLY.
Remedial action
- Replace