Product problems

  • Device Alarm System
  • High Test Results

Patient

YR

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE ABNORMALLY HIGH PO2 VALUES ARE CLOSE TO THE VALUES OF ATMOSPHERIC AIR WHICH SUGGESTS, THAT AIR-BUBBLES HAS BEEN INTRODUCED INTO THE SAMPLES, AND THAT THE MEASUREMENTS MAY HAVE BEEN CONDUCTED ON THE SECLUDED AIR.

Description of Event or Problem

WHEN MEASURING PO2 ON THE BLOOD GAS ANALYZER WITH FLEXQ, A PHYSICIAN CONSIDERED THAT THE OBTAINED PO2 VALUES TO BE ABNORMALLY HIGH. THIS WAS CONFIRMED BY NEW MEASUREMENTS. NO ALARM WAS ISSUED. THE VALUES WERE CLOSE TO THE VALUES OF ATMOSPHERIC AIR WERE OBTAINED. THERE HAS BEEN NO INFORMATION OF SERIOUS INJURY IN THE REPORTED INFORMATION.