MTS 24 PLACE CENTRIFUGE

Received Jun 4, 2007 · Event occurred May 9, 2007

Report 1056600-2007-00156 · MDR key 859217

Device

Generic name

Centrifuge

Model number

NA

Catalog number

MTS515060

Lot number

NA

Product problems

  • Material Frayed

Patient

YR

  • No Patient Involvement

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE FRAYED INNER GREEN WIRING OF THE RETRACTILE CABLE OF THE MTS CENTRIFUGE WAS EXPOSED. THE CUSTOMER INDICATED THAT THE ISSUE WAS DISCOVERED DURING ROUTINE YEARLY MAINTENANCE. AN EXPOSED OR BROKEN WIRE MAY HAVE THE REMOTE POTENTIAL FOR ELECTRICAL SHOCK OR FIRE.

Additional Manufacturer Narrative

THE CUSTOMER WAS INSTRUCTED BY OCD CUSTOMER TECHNICAL SVS (CTS) TO DISCONTINUE THE USE OF THE INSTRUMENT. THE REPLACEMENT RETRACTILE CABLE PART NUMBER WAS PROVIDED TO THE CUSTOMER. THIS CUSTOMER HAS NOT LOGGED ANY COMPLAINTS AGAINST THIS INSTRUMENT SINCE THIS INCIDENT. PROD LABELING CAUTIONS THE USER OF THE RISK OF ELECTRIC SHOCK INVOLVED IN SERVICING THE INSTRUMENT. PROD LABELING PROVIDES THE USER WITH THE PROPER PROCEDURE TO FOLLOW WHEN REPAIRS ARE NECESSARY. THE ROOT CAUSE FOR THE DAMAGED WIRE IS UNK.