OTTOBOCK HARMONY® P4 HD MECHANICAL PUMP
Received Apr 25, 2019 · Event occurred Mar 12, 2019
Report 1721652-2019-00002 · MDR key 8551262
Device
Generic name
4r181 Harmony® P4 Hd Mechanical Pump
Manufacturer
Otto Bock Healthcare LpModel number
4R181Catalog number
4R181
Product problems
- Device Difficult to Setup or Prepare
- Device Difficult to Setup or Prepare
Patient
Not reported
- Joint Dislocation
- Collapse
- Joint Dislocation
- Collapse
Narrative
Description of Event or Problem
AS REPORTED BY THE PRESCRIBING / FITTING PRACTITIONER ((B)(6)), "WE RECENTLY HAD A PATIENT BEING FIT WITH A P4 PUMP. DURING THE TEST FIT PHASE, OUR TECHNICIANS HAD SCRATCHED UP THE TOP OF THE MOUNTING PLATE IN ORDER TO GIVE IT MORE BONDING SURFACE AREA. THE MOUNTING PLATE WAS ATTACHED TO THE SOCKET USING FAB TEC TWO PART RESIN. THEN THE PLATE AND SOCKET WERE BONDED TOGETHER WITH FIBERGLASS WRAP. THE PATIENT LEFT OUR FACILITY AND LATER THAT NIGHT THE MOUNTING PLATE COMPLETELY SEPARATED FROM THE SOCKET. UNFORTUNATELY, THE PATIENT FELL AND DISLOCATED HIS SHOULDER."
Remedial action
- Other