inforMED
MalfunctionJSD

MICROSCAN PROMPT INOCULATION SYSTEM-D

Received Oct 9, 2006 · Event occurred Sep 28, 2006

Report 2919016-2006-00038 · MDR key 854682

Device

Generic name

Panel Inoculating Supplies

Manufacturer

Dade Behring, Inc.

Model number

NA

Catalog number

B1026-10D

Lot number

20071127

Product problems

  • No Display/Image

Patient

YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

ACCOUNT REPORTED TO DADE BEHRING THAT THEY HAD INTERMITTENTLY OBSERVED INSUFFICIENT GROWTH (NO RESULTS) ON QC AND CLINICAL ISOLATES. ISSUE WAS ONLY ASSOCIATED WITH THE IDENTIFIED PROMPT LOT. ACCEPTABLE RESULTS WERE OBTAINED AFTER REPEATING THE TESTS. THERE WERE NO REPORTS OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.

Additional Manufacturer Narrative

PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMERS OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.

Remedial action

  • Other