MICROSCAN PROMPT INOCULATION SYSTEM-D
Received Oct 9, 2006 · Event occurred Sep 28, 2006
Report 2919016-2006-00037 · MDR key 854624
Device
Generic name
Panel Inoculating Supplies
Manufacturer
Dade Behring, Inc.Model number
NACatalog number
B1026-10D
Lot number
20071127
Product problems
- No Display/Image
- Device Displays Incorrect Message
Patient
YR
- Brain Injury
Narrative
Description of Event or Problem
ACCOUNT REPORTED TO DADE BEHRING THAT THEY HAD OBSERVED SOME INSUFFICIENT GROWTH (NO RESULTS) THAT CLEARED UPON RETEST. REPORT WAS ASSOCIATED WITH ONLY THE IDENTIFIED PROMPT LOT AND ERRONEOUS REPORTS WERE NOT SENT OUT TO THE PHYSICIAN. NO REPORT OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.
Additional Manufacturer Narrative
PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMERS OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.
Remedial action
- Other