MICROSCAN PROMPT INOCULATION SYSTEM-D

Received Oct 9, 2006 · Event occurred Sep 28, 2006

Report 2919016-2006-00037 · MDR key 854624

Device

Generic name

Panel Inoculating Supplies

Manufacturer

Dade Behring, Inc.

Model number

NA

Catalog number

B1026-10D

Lot number

20071127

Product problems

  • No Display/Image
  • Device Displays Incorrect Message

Patient

YR

  • Brain Injury

Narrative

Description of Event or Problem

ACCOUNT REPORTED TO DADE BEHRING THAT THEY HAD OBSERVED SOME INSUFFICIENT GROWTH (NO RESULTS) THAT CLEARED UPON RETEST. REPORT WAS ASSOCIATED WITH ONLY THE IDENTIFIED PROMPT LOT AND ERRONEOUS REPORTS WERE NOT SENT OUT TO THE PHYSICIAN. NO REPORT OF INJURY OR ILLNESS ASSOCIATED WITH THIS ISSUE.

Additional Manufacturer Narrative

PERFORMANCE TESTING OF CUSTOMER RETURNS AND RETENTION SAMPLES. IN-HOUSE TESTING PERFORMED AND CONFIRMED INSUFFICIENT GROWTH ISSUE WITH CUSTOMER PROVIDED SAMPLES AND ALSO WITH RETENTION SAMPLES. DADE BEHRING HAS INITIATED A FIELD CORRECTION FOR THIS ISSUE AND NOTIFIED CUSTOMERS OF THE POTENTIAL FOR DISCREPANT MIC AND/OR IDENTIFICATION RESULTS WITH CLINICAL AND QC ISOLATES WITH THE IDENTIFIED PROMPT INOCULATION SYSTEM-D LOT.

Remedial action

  • Other