inforMED
InjuryEXB

PREMIER FLEXWEAR DRAINABLE POUCH WITH FILTER AND TAPE BORDER

Received Apr 16, 2019 · Event occurred Mar 8, 2019

Report 1119193-2019-00011 · MDR key 8518471

Device

Generic name

Premier Flexware Drainable Pouch With Filter And Tape Border

Catalog number

8081

Lot number

9A132

Product problems

  • Appropriate Device Problem Term/Code Not Available
  • Appropriate Device Problem Term/Code Not Available

Patient

66 YR

  • Hypersensitivity/Allergic reaction
  • Hypersensitivity/Allergic reaction

Narrative

Additional Manufacturer Narrative

BIOCOMPATIBILITY TESTING HAS BEEN PERFORMED IN ACCORDANCE WITH ISO 10993 AND MATERIALS WERE FOUND TO BE NON-SENSITIZERS. DHR REVIEW CONDUCTED BASED UPON THE LOT NUMBER PROVIDED AND NO ISSUES IDENTIFIED. TREND DATA REVIEWED AND NO TREND OBSERVED. THE ROOT CAUSE OF THE END USERS GENERALIZED RASH CANNOT BE DETERMINED.

Description of Event or Problem

IT WAS REPORTED THAT AN END USER EXPERIENCED A RASH OVER HIS WHOLE BODY AND ITCHING ASSOCIATED WITH THE USE OF THE OSTOMY APPLIANCE. END USER WAS PRESCRIBED HYDROXYZINE PILLS AND INSTRUCTED TO USE HYDROCORTISONE CREAM. HE STATES IT HELPS A LITTLE.