MalfunctionCAN
AMVEX MINI CLICK STYLE GAS REGULATOR
Received Apr 8, 2019
Report MW5085650 · MDR key 8495914
Device
Generic name
Portable Oxygen Regulator / Regulator Vacuum
Manufacturer
Ohio Medical LlcModel number
EB-GPM2-870RM-TProduct problems
- Loose or Intermittent Connection
- Increase in Pressure
- Improper Flow or Infusion
- Separation Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
DURING NORMAL USE, DEVICE SUDDENLY STARTED FLOWING HIGH PRESSURE OXYGEN OUT OF THE PRESSURE RELIEF HOLE. THIS IS ONE OF MULTIPLE REPORTS YOU WILL RECEIVE FOR THE SAME MODEL (DIFFERENT SERIAL NUMBERS) DEVICE WITH EXACTLY THE SAME FAILURE. MFR WAS NOTIFIED MULTIPLE TIMES WITH ACKNOWLEDGEMENT THAT A PROBLEM EXISTS. REPLACEMENT UNITS ALSO FAILED. DEVICE CLICK KNOB ALSO WAS NOT SECURE. ROUTINELY CAME OFF OF DEVICE. OXYGEN TUBING NIPPLE ALSO NOT SECURE. WE CONSTANTLY EXPERIENCED THE NIPPLE SEPARATING FROM THE DEVICE.