CLINITEK STATUS+

Received Apr 2, 2019 · Event occurred Feb 17, 2019

Report 3002637618-2019-00038 · MDR key 8475522

Device

Generic name

Clinitek Status+

Catalog number

10379679

Product problems

  • False Negative Result
  • False Negative Result

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE CUSTOMER STATED THAT REPEAT TESTING WAS PERFORMED TO CONFIRM CORRECT RESULTS WHICH MET THE CLINICAL PICTURE. SIEMENS HAS REQUESTED THAT THE REAGENT BE RETURNED FOR INVESTIGATION AND THE CUSTOMER HAS AGREED. THERE WERE NO ISSUES WITH TECHNIQUE OR MAINTENANCE FOUND. THE CAUSE OF THIS EVENT IS UNKNOWN.

Description of Event or Problem

THE CUSTOMER REPORTED FALSE NEGATIVE LEUKOCYTE RESULTS ON SEVERAL PATIENTS ON THE CLINITEK STATUS+ COMPARED TO THE MICROSCOPIC EXAMINATION OF THE SEDIMENT. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

Additional Manufacturer Narrative

UPDATED CONTACT OFFICE - NAME/ADDRESS. SIEMENS TECH OPS HAS COMPLETED THE INVESTIGATION: THE CUSTOMER SUPPLIED THE REAGENT - MULTISTIX 10 SG (7 STRIPS), BUT NO PATIENT SAMPLES WERE RECEIVED. THE STRIPS WERE VISUALLY INSPECTED FOR ABNORMALITIES, NONE WERE FOUND. THE DESSICANT WAS IN THE BOTTLE AS IT SHOULD BE. USING THE CUSTOMERS STRIPS, THEY WERE DIPPED INTO LEUKOCYTE ESTERASE RELEASE SOLUTION - 10 CELLS/DL AND RUN ON THE STATUS+. SEVEN TRACE LEUKOCYTE RESULTS WERE OBTAINED. THERE WERE NO FALSE NEGATIVE RESULTS OBSERVED. THE CAUSE OF THE EVENT IS UNKNOWN. PER THE CUSTOMER BULLETIN "CAUSES OF LEUKOCYTE DISCREPANCIES BETWEEN REAGENT STRIP AND MICROSCOPIC METHOD", THE CELL MEMBRANES OF LEUKOCYTES TEND TO LYSE MORE READILY IN URINES WITH LOWER SPECIFIC GRAVITY (HYPOTONIC), BUT THEY REMAIN INTACT FOR LONGER PERIODS IN URINES WITH HIGHER SPECIFIC GRAVITY (HYPERTONIC). SINCE THE REAGENT STRIP REACTS WITH THE LEUKOCYTE ELASTASE (AN ESTERASE) CONTAINED WITHIN THE CELL, IT IS NECESSARY FOR SOME OF THE CELLS TO HAVE LYSED IN ORDER TO BE DETECTED WITH THE REAGENT STRIP. THEREFORE, URINE SAMPLES WITH HIGHER SPECIFIC GRAVITIES TEND TO HAVE FEWER LYSED CELLS, WHICH WOULD PRODUCE FEWER POSITIVE REAGENT STRIP RESULTS. MICROSCOPICALLY, THE INTACT LEUKOCYTES WOULD STILL BE VISIBLE. THE CAUSE OF THE EVENT IS UNKNOWN.