inforMED
MalfunctionIOY

SUPPORT ARM 178

Received Mar 8, 2019 · Event occurred Feb 28, 2019

Report 8010042-2019-00147 · MDR key 8404259

Device

Generic name

Support, Arm

Catalog number

6693715

Product problems

  • Detachment of Device or Device Component

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

IT WAS NOTED THAT THE RAIL CLAMPS DOUBLE PUSH BUTTON DESIGN WITH THE 2 BALL BEARINGS TENDS TO GET STUCK IN THE ADJUSTMENT POSITION AND DON¿T ALWAYS RELEASE. THIS HAS MOST LIKELY CONTRIBUTED TO THE DISENGAGING OF THE SUPPORT ARM FROM THE VENTILATOR CARRIER. THE ROOT CAUSE WHY THE REPORTED SUPPORT ARMS LEVER ACTUATOR GOT STUCK HAS NOT BEEN ESTABLISHED DUE TO THE LACK OF RETURNED PART. WORK ON AN IMPROVED RAIL CLAMP IS CURRENTLY UNDERWAY BY BY OUR RESEARCH AND DEVELOPMENT DEPARTMENT.

Description of Event or Problem

IMPORTER REF. #:(B)(4). MANUFACTURER REF. #: 1909MCC0859.

Additional Manufacturer Narrative

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC, 45 BARBOUR POND DRIVE, WAYNE, NJ 07470. CONTACT PERON: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM FOR PATIENT TUBES FELL OF THE SIDE RAIL OF THE VENTILATOR. THERE WAS NO PATIENT INVOLVEMENT. (B)(4).